Transjugular liver biopsy: modified Ross transseptal needle versus quick-core biopsy needle

Abdom Imaging. 2000 Sep-Oct;25(5):483-5. doi: 10.1007/s002610000043.

Abstract

Background: We evaluated the differences in sample adequacy and safety between a transseptal needle and Quick-core biopsy needle for transjugular liver biopsy.

Methods: Eighteen consecutive patients who had a bleeding diatheses and/or ascites underwent transjugular liver biopsy using a transseptal needle (11 patients) and Quick-core biopsy needle (seven patients). The length of the specimens was measured before fixation. A pathologist reviewed histologic slides for sample adequacy and pathologic diagnoses. Clinical records were reviewed for complication.

Results: In all patients, liver biopsy was successful. A total of 45 specimens were obtained, with an average of 2.5 passes per patient. The length of specimen was significantly longer with the Quick-core biopsy needle than with the transseptal needle (p<0.05). Biopsied tissue was fragmented in 17 of 25 specimens with the transseptal needle but not fragmented in any specimen with the Quick-core biopsy needle. All specimens were determined to be adequate except one with the transseptal needle. There was no early or delayed complication in any patient.

Conclusion: Transjugular liver biopsy is a safe and effective procedure without any significant difference in complication and adequacy when using a transseptal needle or Quick-core biopsy needle. Larger specimens can be obtained without tissue fragmentation with the Quick-core biopsy needle.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Biopsy, Needle / methods*
  • Catheterization, Peripheral*
  • Equipment Design
  • Female
  • Humans
  • Jugular Veins
  • Liver Diseases / pathology*
  • Male
  • Middle Aged
  • Needles*
  • Reproducibility of Results