Article Text

Download PDFPDF

Original Article
A comparative effectiveness trial of two faecal immunochemical tests for haemoglobin (FIT). Assessment of test performance and adherence in a single round of a population-based screening programme for colorectal cancer
  1. Basilio Passamonti1,
  2. Morena Malaspina1,
  3. Callum G Fraser2,
  4. Beatrice Tintori1,
  5. Angela Carlani1,
  6. Valentina D'Angelo1,
  7. Paola Galeazzi1,
  8. Eugenio Di Dato1,
  9. Loretta Mariotti1,
  10. Simonetta Bulletti1,
  11. Maria Rosaria D'Amico1,
  12. Daniela Gustinucci1,
  13. Nadia Martinelli1,
  14. Nicoletta Spita1,
  15. Elena Cesarini1,
  16. Tiziana Rubeca3,
  17. Mariadonata Giaimo4,
  18. Nereo Segnan5,
  19. Carlo Senore5
  1. 1Azienda USL Umbria 1 Perugia, Laboratorio Unico di Screening, Perugia, Italy
  2. 2Centre for Research into Cancer Prevention and Screening, University of Dundee School of Medicine, Dundee, UK
  3. 3Laboratorio Regionale Prevenzione Oncologica ISPO-Istituto per lo Studio e la Prevenzione Oncologica, Firenze, Italy
  4. 4Servizio Prevenzione, Sanità Veterinaria, Direzione Regionale Salute, Perugia, Italy
  5. 5Città della Salute e della Scienza University Hospital, SC Epidemiology, Screening, Cancer Registry, CPO, Turin, Italy
  1. Correspondence to Morena Malaspina, Azienda USL Umbria 1, Laboratorio Unico di Screening, Via XIV Settembre n. 75 Perugia, Italy; morena.malaspina{at}uslumbria1.it

Abstract

Aim To compare acceptability and diagnostic accuracy of a recently available faecal immunochemical test (FIT) system (HM-JACKarc) with the FIT routinely used in an established screening programme (OC-Sensor).

Design Randomised controlled trial (ISRCTN20086618) within a population-based colorectal cancer (CRC) screening programme. Subjects eligible for invitation in the Umbria Region (Italy) programme were randomised (ratio 1:1) to be screened using one of the FIT systems.

Results Screening uptake among the 48 888 invitees was the same for both systems among subjects invited in the first round and higher with OC-Sensor than with HM-JACKarc (relative risk (RR): 1.03; 95% CI 1.02 to 1.04) among those invited in subsequent rounds. Positivity rate (PR) was similar with OC-Sensor (6.5%) as with HM-JACKarc (6.2%) among subjects performing their first FIT screening and higher with OC-Sensor (5.6%, RR: 1.25, 95% CI 1.12 to 1.40) than with HM-JACKarc (4.4%) among those screened in previous rounds. Positive predictive value (PPV) (OC-Sensor: 25.9%, HM-JACKarc: 25.6%) and detection rate (DR) (OC-Sensor: 1.40%; HM-JACKarc: 1.42%) for advanced neoplasia (AN: CRC + advanced adenoma) were similar among subjects performing their first FIT screening. The differences in the AN PPV (OC-Sensor: 20.3%, HM-JACKarc: 22.6%) and DR (OC-Sensor: 0.96%, HM-JACKarc: 0.83%) among those screened in previous rounds were not statistically significant. The number needed to scope to detect one AN was 3.9 (95% CI 5.8 to 2.9) and 3.9 (95% CI 5.5 to 2.9) at first and 4.9 (95% CI 5.8 to 4.2) and 4.4 (95% CI 5.3 to 3.7) at subsequent screening, with OC-Sensor and HM-JACKarc, respectively.

Conclusions Our results suggest that acceptability and diagnostic performance of HM-JACKarc and of OC-Sensor systems are similar in a screening setting.

Trial registration number ISRCTN20086618; Results.

  • CANCER PREVENTION
  • SCREENING
  • COLORECTAL CANCER SCREENING

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Significance of this study

What is already known on this subject?

  • Faecal immunochemical tests (FIT) for haemoglobin (Hb) have been recommended as the preferred method for population colorectal cancer screening.

  • Several different FIT have been introduced and standardisation on faecal Hb concentration has been recommended for comparison.

  • Available comparative information about diagnostic performance of different methods is mainly derived through laboratory-based studies.

What are the new findings?

  • The acceptability and the diagnostic performance of two FIT systems (OC-Sensor and HM-JACKarc) are similar in a screening setting.

  • Diagnostic performance of FIT systems may be influenced by subject's screening history (ie, expected prevalence of the disease).

  • Setting the faecal Hb cut-off to achieve the same positivity rate with the two systems resulted in the same advanced neoplasia yield and positive predictive value.

How might it impact on clinical practice in the foreseeable future?

  • Adopting faecal Hb concentration cut-offs to give equivalent positivity rates, thus using the same endoscopy resources, allows for a direct comparison of the performance of different FIT. This strategy might reduce the variability related to the individual FIT methodology, since this could alter the relationship between Hb concentration and positivity rate.

Introduction

Following the recommendation for Europe to adopt faecal immunochemical tests (FIT) for haemoglobin (Hb) as the initial non-invasive investigation of choice for population-based colorectal cancer (CRC) screening programmes,1 several different qualitative and quantitative FIT have become available for use in both screening and clinical settings. While the availability of several tests may stimulate competition, lower costs and promote technical improvements, data concerning the diagnostic performance of FIT are still mainly derived from laboratory-based evaluations2–3 from studies conducted in clinical settings4–6 or from studies using a paired design.7–9 The availability of reliable information about the performance of new FIT in screening settings is crucial, however, since even relatively small variations in performance characteristics can have substantial impacts on programme organisation and delivery.

The positivity rate (PR) that personnel and financial resources can accommodate varies across regions and limited endoscopy capacity has been identified as a barrier to initiation and expansion of screening in several settings.10–11 An increase in the number of colonoscopes resulting from a higher PR could lead, for example, to longer waiting times for assessment of participants with positive test results and this could eventually lead to a reduction in the population invited for screening. Similarly, a reduction in the positive predictive value (PPV) can negatively affect the cost-benefit balance of a programme. Moreover, practical aspects, for example, the design and ease of use of the sample collection device can play a role when the selected population is invited to participate.2 ,11 The device should be designed such that inadequate sampling or leakage are minimised. It should be easy to label with adequate information to identify the sample and it should also be suitable for packing and dispatch by normal post. The users' preferences for different devices therefore need to be assessed. Finally, the costs of the system and of the postage, when relevant, need to be assessed as well.11

Comparative effectiveness studies, conducted in the context of established CRC screening programmes, as performed in a recent study from The Netherlands,12 may therefore represent a valuable methodology to gather the information relevant to assist screening planning and decision-making about the adoption of screening tests. We conducted a randomised controlled trial (RCT) within a population-based CRC screening programme to compare the acceptability and the diagnostic accuracy the FIT routinely used in a well-established programme (OC-Sensor, Eiken Chemical, Tokyo, Japan) with a recently available automated quantitative FIT system (HM-JACKarc, Kyowa Medex, Tokyo, Japan).

Methods

The Umbria Region CRC screening programme

An organised population-based CRC screening programme was started in the Umbria Region (Italy) in 2006. All residents aged 50–74 years are targeted for invitation biennially and asked to collect a single faecal sample for subsequent analysis by FIT. The FIT sample collection device is mailed to the potential participants, together with a personal invitation letter signed by the appropriate general practitioner (GP), an information leaflet, the instructions for collecting and storing the sample and a prepaid envelope for returning the device. Each device is identified by the participant's name and social security number in bar-code format. The handling of the devices, as well as the test results, is fully automated: participant's data are entered in the screening database using a bar-code reader by the laboratory staff when the device is processed and the results of the analytical determination of the faecal haemoglobin concentration (f-Hb) are directly recorded in the same database, linked to the analytical system.

Subjects with a positive test result are invited by mail for a colonoscopy in one of five reference endoscopy centres, which reserve dedicated sessions for participants in the programme. The entire screening process, and treatment when necessary, are offered free of charge, following the quality standards recommended by national13 and European Union guidelines.1 An automated linkage between the screening and pathology databases supports timely and comprehensive recording of all colonoscopy and treatment results for each participant with a positive test result.

Randomisation and invitation

We designed an RCT (ISRCTN20086618) within our regional programme; the study was approved by the local Ethics Review Board (Prot. n.3032/14/AV 29/04/2014). Subjects eligible for invitation during the enrolment period (from 6 November 2014 to 31 March 2015) were randomly allocated (ratio 1:1) to be screened with one of the two FIT systems, namely, OC-Sensor and HM-JACKarc. The randomisation was stratified by gender, age group (five 5-year age groups between 50 and 74 years of age), screening history (first and subsequent screening) and area of residence (urban and rural). The scheme was computer generated within the screening programme IT system and identified when individuals were to be invited to participate. The process of generating and mailing the different invitation materials was fully automated, therefore, blinding the researchers to the allocation of the intervention to individuals.

Following the standard screening protocol, potential participants with a previous diagnosis of CRC (information derived from the hospital discharge records database), those having performed a FIT within the previous 2 years or those who had a colonoscopy within the previous 5 years (identified from information available from the regional screening and endoscopy databases) and those suffering from terminal illness, or unable to provide informed consent (GP indication), were excluded from the invitation. A reminder letter was mailed to all non-responders 3 months following the initial invitation. The invitation kit mailed for the study included an informed consent form and a leaflet explaining the design and the rationale for the study. The invitation, containing pictorial and written materials, was designed to be as similar as possible for each of the analytical systems.

Participants were instructed to store the completed device at 4°C and to send it back to the central laboratory as soon as possible. Devices were collected and forwarded to the screening laboratory every day by the postal company, following the routine procedures of the screening programme.

Laboratory and quality control procedures

Following the Faecal Immunochemical Tests Evaluation Reporting guidelines, f-Hb results were reported as µg Hb/g faeces, as recommended by the Expert Working Group on FIT for Screening, the Colorectal Cancer Screening Committee, World Endoscopy Organisation.14 All analyses were performed in the central laboratory of the Umbria screening programme (National accreditation: CERMET ISO 9000:2000, ISO 9001:2008, ISO 9004:2000, UNI EN 19011:2003; D.D. n. 9628 del 21/11/2014) by three experienced laboratory medicine professionals. A detailed description of the calibration procedures and the observed analytical coefficients of variation for the two systems is available in online supplement 1.

The OC-Sensor device collects 10 mg faeces with a serrated probe attached to the cap into 2.0 mL buffer.2 The cap is pushed on to the device which incorporates a small filter that removes faecal particulate matter from the sample in buffer before analysis, a feature that reduces the possibility of incorrect sampling. The analytical working range is 10–200 µg Hb/g faeces. Dilution is semi-automated after manual identification of samples with f-Hb higher than the upper analytical limit and manual replacement of the sample tube on the analyser. This dilution extends the analytical range beyond the undiluted upper analytical limit of 500 μg Hb/g faeces.2 The HM-JACKarc device collects 2 mg of faecal material in dimples in the probe attached to the cap of the device into 2.0 mL buffer; the cap is screwed onto the sampling device. The analytical working range is 7–400 µg Hb/g faeces: the system has no automated system for automatic dilution and re-analysis of samples that have an f-Hb higher than the upper analytical limit.2

Faecal samples were stored at 4°C in the laboratory until the time of the analysis, which was usually performed on the day of receipt and no later than the following day. An identical f-Hb cut-off criterion for referral for colonoscopy (20 µg Hb/g faeces) was used for both systems. The f-Hb results were recorded electronically and were automatically linked to the participant's record in the screening database.

Colonoscopy

Colonoscopy was performed in five regional reference centres using standard high-definition endoscopes following a cathartic bowel preparation. The examination was considered complete if the caecum was visualised or, in the case of failure, when a subsequent colonoscopy, performed within the following 6 months, reached the caecum; the combined results of the two examinations were considered for analysis. All detected lesions were measured with open biopsy forceps and annotated according to size, morphology and localisation. Histology was defined according to the WHO criteria15: advanced adenoma (AA) was defined as an adenoma with any of the following features: size ≥10 mm, high-grade dysplasia or villous component >20%; cancer was defined as the invasion of malignant cells beyond the muscularis mucosae.

Statistical analysis

The performance indicators considered in our comparative analysis included proportion of inadequate tests, positivity rate (PR), PPV, detection rate (DR) for CRC and AA, number needed to scope (NNScope) to detect one advanced neoplasm (AN=CRC+AA). We also compared the participation rate with the two methods, although there was no reason to expect major differences in uptake given that the two systems do not show substantial differences when characteristics such as the design and handling of the specimen collection devices possibly affecting their acceptability are considered. However, it might be that offering a new device, albeit only slightly different from that promoted in the mass-media campaigns encouraging participation, might have some impact on the uptake.

Sample size was determined based on the expected participation rate (50%), PR (5%) and AN DR (1%), derived from the observed results of our screening programme. Based on these assumptions, enrolling 25 000 participants for each analytical system would allow detection of a 1.0%, 0.4% and 7.0% absolute difference between the two systems in positivity, AN DR and PPV, respectively, as statistically significant (α: 0.05; power: 80%). The exact method was used to calculate the 95% CIs of proportions; 95% CI for the NNScope were calculated as the inverse of the 95% CI of the corresponding PPV. Relative risk (RR) with 95% CI was used as a measure of association of the outcomes of interest between analytical systems, setting the OC-Sensor as the reference system. The standardised estimates were compared using a direct standardisation method assuming the total population included in the study as the reference for calculations. All statistical tests were two-sided and a p value of <0.05 was considered to be statistically significant.

Results

During the recruitment period, 48 888 people (22 840 men and 26 048 women) were invited: 47.2% of them were living in urban areas (>150 subjects/km2) and 52.8% in rural areas. The participation was 57.4% in the former and 63.2% in the latter. Out of the 48 888 invitees, 24 314 (11 319 men and 12 995 women) were randomised to the OC-Sensor group and 24 574 (11 521 men and 13 053 women) to the HM-JACKarc group. The two groups were comparable in age, gender and screening history (table 1).

Table 1

Characteristics of participants by FIT system, gender, age and screening history

Participation (table 2) was the same for both systems among individuals invited for the first time in the programme (first screening round), while it was slightly higher with OC-Sensor than with HM-JACKarc (RR: 1.03; 95% CI 1.02 to 1.04) for those who had already been invited in previous rounds in the programme (subsequent screening rounds). A similar trend towards a lower participation among men than among women invited in the first screening round was observed both with OC-Sensor (RR: 0.80 (95% CI 0.73 to 0.88)) and with HM-JACKarc (RR: 0.83 (95% CI 0.76 to 0.92)); an opposite trend was observed among people having already participated in previous rounds, showing a higher participation among men than among women (OC-Sensor—RR: 1.04 (95% CI 1.02 to 1.05); HM-JACKarc—RR: 1.02 (95% CI 1.01 to 1.04)). Participation was higher in the youngest age group (50–54 years) for both systems at first screening and among people aged 60–69 years as compared with younger or older age groups at subsequent screening rounds.

Table 2

Uptake of the screening invitation with the OC-Sensor and HM-JACKarc systems by gender, age and screening history

The overall PR (table 3) was 6.4% (271/4247) at first and 5.0% (1240/24 751) at subsequent screening and, as expected, it was higher among men than among women and increased with age quintile. Positivity was similar with OC-Sensor (6.5%) as with HM-JACKarc (6.2%, RR: 1.05, 95% CI 0.84 to 1.33) at first screening, while it was higher with OC-Sensor (5.6%, RR: 1.25, 95% CI 1.12 to 1.40) than with HM-JACKarc (4.4%) at subsequent screening, with a similar trend among men and women.

Table 3

Positivity and colonoscopy performed and completed for OC-Sensor and HM-JACKarc systems by gender, age and screening history

The uptake of colonoscopy in those with positive test results was the same (table 3) with OC-Sensor (overall: 84.1%, men: 85.7%, women: 82.5%) as with HM-JACKarc (overall: 83.8%, men: 83.3%, women: 84.3%), as was the completion rate of the examinations. Out of 672 subjects who undertook an assessment colonoscopy in the OC-Sensor group, 29 (4.1%) repeated the investigation within 6 months, following an incomplete index colonoscopy (N=12), or to check completeness of polypectomy; similarly, in the HM-JACKarc group, 15 out of 553 (2.6%; p=0.166) repeated the investigation within 6 months (2 following inadequate preparation and 12 to assess the site of polypectomy).

PPV for AN was similar for the two systems (OC-Sensor: 25.9%, HM-JACKarc: 25.6%) at first screening (table 4) as it was also for the AN DR (OC-Sensor: 1.40%; HM-JACKarc: 1.42%).

Table 4

Detection rate and PPV for advanced adenoma and CRC by gender, age and screening history

However, the PPV for AN at subsequent screening of a positive test result was lower with OC-Sensor (20.3%; 95% CI 17.4% to 24.1%) than with HM-JACKarc (22.6%, 95% CI 18.9% to 26.8%), but the difference did not reach the level of statistical significance. The AN DR was slightly higher, but not statistically significant, with OC-Sensor (0.96%) than with HM-JACKarc (0.83%). The NNScope to detect one case of AN was 3.9 (95% CI 5.8 to 2.9) and 3.9 (95% CI 5.5 to 2.9) at first screening with OC-Sensor and HM-JACKarc, respectively; the corresponding figures at subsequent screening were 4.9 (95% CI 5.8 to 4.2) and 4.4 (95% CI 5.3 to 3.7).

Staging was available for all CRC detected at first screening both with OC-Sensor (N=5; UICC stage: I=4; II=1) and with HM-JACKarc (N=5; UICC stage: I=2; II=1; III=2); staging information for CRC detected at subsequent screening was available for 11 (UICC stage: I=7; II=1; III=3) out of 14 cases detected with OC-Sensor and for 15 (UICC stage: I=9; II=1; III=5) out of 16 cases detected with HM-JACKarc. The proportion of CRC located in the proximal (to the descending colon) segments was 26.3% and 27.8% with OC-Sensor and HM-JACKarc, respectively; the corresponding data for AN were 41.9% and 33.0%. Neither difference reached statistical significance.

The results remained the same when considering standardised comparisons (data not shown): the PR was slightly different among participants at the initial screening test (OC-Sensor: 6.3% and HM-JACKarc: 6.6%), but the difference was not statistically significant; the crude and standardised estimates for PR, PPV and DR were the same when considering participants in subsequent screening rounds.

After increasing the f-Hb cut-off criterion for PR of OC-Sensor at subsequent screening (from 20 to 29 µg Hb/g faeces) to achieve the same PR as observed with HM-JACKarc (data not shown), the difference in the AN yield between the two methods was substantially reduced (OC-Sensor: PPV: 21.0%; DR: 0.78%; NNScope: 4.8).

Discussion

In an RCT undertaken in the context of a population-based CRC screening programme, OC-Sensor and HM-JACKarc FIT analytical systems showed similar participation patterns by gender, age and screening history. The observed 3.0% increase in the participation rate with OC-Sensor among individuals having already participated in previous rounds may be related to the initial impact of a new and slightly different specimen collection device. Since all invitation kits were sent by post, following the well-established procedures of the programme, individuals already used to perform the faecal sampling with the previous device might have faced difficulties in understanding the reasons for change, possibly enhancing negative attitudes to the experimental context. Importantly, those invited for their first time in the programme showed the same response with both systems, which would support this hypothesis. The lower participation found among men when compared with women across all age groups, as well as the trends for age and screening history are consistent with the results in Italian screening programmes.16

PR, PPV and AN DR among participants undertaking their first screening were similar, while a higher PR among people with previous negative test results was observed when using the OC-Sensor as compared with the HM-JACKarc. Since the majority of participants screened after completion of the roll-out phase of a screening programme are expected to have already taken part in previous rounds (about 85% in the current programme), the reduction in the PR in this group will have a favourable impact on the endoscopy workload of a population-based programme. The PPV for AN was also slightly decreased with OC-Sensor as compared with the HM-JACKarc, when considering subjects with previous negative test results, but the difference did not reach statistical significance, since the DR was similar for CRC, but slightly higher for AA with the OC-Sensor than with HM-JACKarc. Although the observed increase was not statistically significant, it must be realised that our study was powered to detect larger differences. Assuming that the observed increase could be declared statistically significant in a larger sample, it can be estimated that about eight additional colonoscopes would be necessary to detect one additional AN over 1000 participants with OC-Sensor. Whether such an increase in the colonoscopy workload to achieve this small increase in the DR would be acceptable and sustainable can likely be determined only for each programme individually. Availability of such information can, however, inform a more explicit decision-making process.

Following the proposal to standardise the measuring units of f-Hb to µg Hb/g faeces,14 to take into account the different masses of faeces collected and the volumes of the buffer in the devices, identical f-Hb cut-offs were adopted in for both systems. While similar PR and PPV were observed when using the same f-Hb cut-off among participants performing their first screening test, harmonisation of f-Hb cut-off did not result in similar PR with the two systems among subjects. However, setting a higher f-Hb cut-off criterion for the PR with the OC-Sensor at subsequent rounds, to achieve the same PR with the two systems (29 µg Hb/g faeces rather than 20 µg Hb/g faeces) would result in a similar diagnostic performance. Grobbee et al12 have pointed out recently that comparison between FIT systems is usually done at the same f-Hb. Our findings support their thesis that improved comparison on diagnostic yield would be achieved with PR rather than f-Hb.

A future publication will explore, in detail, a comparative analysis of f-Hb with the two systems, stratified by age, gender, screening history and type of lesions to test this hypothesis. The reasons for the observed difference in the PR between the two systems at subsequent but not at first round, when using the same standardised f-Hb cut-off need to be investigated. The lower prevalence of neoplasia among people with previous negative results might result in lower average f-Hb and, for a number of reasons,12 different methods might show different performances.

The site distribution of the screen-detected lesions observed is consistent with previous reports, reflecting both the expected distribution of CRC and adenoma across colonic site, given the age distribution of our population17 and the documented higher sensitivity of FIT for distal lesions.18

Conflicting data have been reported from previous studies comparing OC-Sensor and HM-JACKarc systems. In a recent Italian study19 inviting 5044 subjects to receive both OC-Sensor and HM-JACKarc devices, after f-Hb cut-off optimisation, OC-Sensor showed a lower PR (3.5% vs 6.2%) and a lower relative sensitivity for AN (73.7% vs 88.2%) than HM-JACKarc. The specificity was 97.6% for the OC-Sensor and 96.0% for the HM-JACK, resulting in a decrease of the required assessments from 5.1% to 3.5% and in an estimated 30% decrease of the costs for every individual investigated for AN. On the other hand, a study comparing two cohorts screened either with OC-Sensor or HM-JACKarc (using the same 20 µg Hb/g faeces cut-off for positivity), in the context of a screening programme in Taiwan,20 showed a similar PR and DR for AA for both systems but a lower CRC DR and a lower interval cancer rate with OC-Sensor than with HM-JACKarc: however, the HM-JACKarc is not an identical system to the HM-JACKarc evaluated in our study. Moreover, differences in the screened populations (including baseline prevalence of disease, characteristics of the screened population with respect to gender, age and screening history) as well as in the f-Hb cut-off criteria for positivity, may explain the observed findings in the Taiwan study.18 These studies do suggest that, even when using a standardised reporting unit system, identical f-Hb may perform differently between systems, while the choice of a suitable f-Hb cut-off for positivity in a population-based screening setting should take into account also the PR, influencing the colonoscopy workload. Our results, showing a similar diagnostic performance of the two systems, may not be directly comparable with the findings from previous reports, since updated systems were used both for OC-Sensor (different buffer) and for HM-JACKarc (different analytical system, collection device and analytical range to the HM-JACKarc system). They would support, however, the conclusion of the Taiwan study suggesting that differences in diagnostic performance may still be expected, despite harmonisation of f-Hb cut-off.

One of the main strengths of our study is that it was conducted in standard care conditions, within an established population-based CRC screening programme, targeting both men and women with different screening histories over the entire screening age range. It can therefore provide relevant indications about the expected impact of the adoption of these systems in an average risk population both at first and subsequent screening. The compliance with colonoscopy referral among subjects with positive screening test results and the colonoscopy completion rate were similar in the two groups: they were higher than the average level of these indicators in Italian programmes and consistent with the results of the regional programme in the more recent years.16 The high quality of the assessment procedures reinforces the validity of our conclusions.

In conclusion, our results would suggest that acceptability and diagnostic performance characteristics of the OC-Sensor and the newer HM-JACKarc systems are similar in a screening setting. The roll-out of programmes and the availability of endoscopy resources should be considered when choosing the f-Hb cut-off criterion for positivity of a FIT system for CRC screening. Selection of the same harmonised f-Hb cut-off may still result in different PR and PPV for AN. Adopting f-Hb cut-offs to give equivalent PR allows instead for a direct comparison of the performance of different systems when using the same amount of endoscopy resources. Using PR as a reference might also reduce the variation possibly related to ongoing technical developments of the specimen collection devices (of the buffer in particular) introduced by manufacturers. A recent study, for example, showed that a change in the formulation of the OC-Sensor device buffer, aimed at stabilising the Hb in the samples and reducing Hb degradation, was associated with higher f-Hb for values below the threshold of 20 μg Hb/g faeces compared with the old buffer.21 Although the difference in the PR observed in that study was not statistically significant, different buffer formulations and other factors might alter the relationship between Hb concentration and PR.

Acknowledgments

CRC screening programme coordinators: Azienda USL Umbria 1, Gabriella Vinti, Azienda USL Umbria 2, Annunziata Di Marco, Rosa Corvetti, Stefania Prandini Servizio Prevenzione, Sanità Veterinaria, Direzione Regionale Salute; endoscopy reference centres: Azienda USL Umbria 1, U.O.C. Gastroenterologia Osp. S. Maria della Misericordia, Perugia e U.O.C. Gastroenterologia Osp. S. Maria, Terni; pathology units: S.S.D. Citologia ed Istologia Diagnostica Osp. S. Maria della Misericordia Perugia e S.C. Anatomia patologica Osp. S. Maria, Terni.

References

Footnotes

  • Contributors BP, MM, MG TR, CS conceived the idea for this work. MM collated the data; CS, MM, CF contributed to the data analysis. CS, MM, CF, NS, wrote drafts of the manuscript. BT, AC, VD'A, PG, EDD performed laboratory tests, LM organised invitations, SB, MRD'A, DG, NM, NS, EC participated in the planning of the study. All authors critically reviewed the submitted manuscript and gave important intellectual input. All authors approved the final version of the manuscript.

  • Funding The study was funded by the Umbria Regional Health Authority Kyowa Medex and Menarini provided the faecal immunochemical test used in the study.

  • Competing interests CGF has undertaken paid consultancy with Immunostics and Kyowa Medex, and received funding for attendance at meetings from Alpha Labs. CS (past 2 years): received support by Medical System for travel and accommodation at a scientific meeting in Italy. His Institution receives PillCam Colon devices from Covidien-Given Imaging to conduct a study and, during 2014, it has received without charge, from EndoChoice, Fuse endoscopy systems to conduct a trial.

  • Patient consent Obtained.

  • Ethics approval The study was approved by the local Ethics Review Board Prot. n.3032/14/AV 29/04/2014.

  • Provenance and peer review Not commissioned; externally peer reviewed.